Figure 8-7. fault 5, open heater -9, Oxymitter 4000, Alarms o2 heater open – Emerson OXYMITTER 4000 User Manual

Page 113: Instruction manual

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0

December 2003

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 8-9

Oxymitter 4000

LOI

KEYPAD

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

37260023

Alarms

O2 Heater Open

Figure 8-7. Fault 5, Open Heater

e. Fault 5, Open Heater

Figure 8-7 shows the electronic assembly
for an Oxymitter 4000 with a membrane
keypad (upper view) and an Oxymitter 4000
with an LOI (lower view).

Membrane Keypad. When Fault 5 is de-
tected, the HEATER LED flashes once,
pauses for three seconds, and repeats.

1. Remove power.

2. Remove the electronic assembly per

paragraph 9-4d, Electronic Assembly
Replacement.

3. Using a multimeter, measure across

the heater connector J8.

4. The measurement should be approxi-

mately 72 ohms. If the heater is open,
see paragraph 9-4h, Heater Strut
Replacement.

LOI. When Fault 5 is detected, the LOI dis-
plays the “O2 Heater Open” message.

1. Remove power. Unscrew and remove

the LOI module from the electronic as-
sembly.

2. Perform the diagnostic steps 2 through

4 shown for the membrane keypad.