Alarms

Very Hi O2 Temp

Figure 8-8. Fault 6, High High Heater Temp

f. Fault 6, High High Heater Temp

Figure 8-8 shows the electronic assembly
for an Oxymitter 4000 with a membrane
keypad (upper view) and an Oxymitter 4000
with an LOI (lower view).

Membrane Keypad. When Fault 6 is de-
tected, the HEATER LED flashes twice,
pauses for three seconds, and repeats.

1. The high high heater temp alarm will

activate when the thermocouple pro-
duces a voltage of 37.1 mV
(1652°F/900°C).

2. The triac and the temperature control

may be at fault.

3. Remove power. Allow the Oxymitter

4000 to cool for five minutes. Restore
power.

4. If the condition repeats, replace the

electronic assembly per paragraph 9-
4d, Electronic Assembly Replacement.

LOI. When Fault 6 is detected, the LOI dis-
plays the “Very Hi O2 Temp” message.

1. The very high O

2

temperature alarm

will activate when the thermocouple
produces a voltage of 37.1 mV
(1652°F/900°C).

2. The triac and the temperature control

may be at fault.

3. Remove power. Allow the Oxymitter

4000 to cool for five minutes. Restore
power.

4. If the condition repeats, replace the

electronic assembly per paragraph
9-4d, Electronic Assembly Replace-
ment.