Troubleshooting 8-15

Oxymitter 4000

Alarms

O2 Cell Bad

LOI

KEYPAD

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

37260029

Figure 8-13. Fault 11, Bad Cell

k. Fault 11, Bad Cell

Figure 8-13 shows the electronic assembly
for an Oxymitter 4000 with a membrane
keypad (upper view) and an Oxymitter 4000
with an LOI (lower view).

Membrane Keypad. When Fault 11 is de-
tected, the O2 CELL flashes three times,
pauses for three seconds, and repeats.

1. The bad cell alarm activates when the

cell exceeds the maximum resistance
value.

2. The cell should be replaced. See para-

graph 9-4i, Cell Replacement, for cell
replacement instructions.

LOI. When Fault 11 is detected, the LOI
displays the “O2 Cell Bad” message. Refer
to the comments and procedures in para-
graphs 1 and 2 above.