Oxymitter 4000 – Emerson OXYMITTER 4000 User Manual

Page 51

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0

December 2003

Rosemount Analytical Inc. A Division of Emerson Process Management

Description and Specifications 1-15

Oxymitter 4000

Self-powered (+5 V), in series with 340 ohms
Conduit ports — 3/4 in.-14 NPT (one threaded hole for
both analog output and logic I/O)

Power Requirements:

Probe Heater ...........................................

175 W nominal

Electronics ...............................................

10 W nominal

Maximum .................................................

500 W

SPS 4000

Mounting ..................................................

Integral to Oxymitter 4000
Remote from Oxymitter 4000

Materials of Construction:

Manifold/Electronics Enclosure ...............

Aluminum

Mounting Brackets ...................................

316 stainless steel (SS)

Pneumatic Fittings ...................................

1/8 in. brass NPT (SS optional)

Pneumatic Tubing ...................................

1/4 in. Teflon (SS optional)

Assembly Hardware ................................

Galvanized and stainless steel

Humidity Range .............................................

100% relative humidity

Temperature Range in Electronics Housing ..

-40

° to 185°F (-40° to 85°C)

Electrical Classification ..................................

NEMA 4X (IP56)

Explosion-Proof Option (both pending) .........

ATEX EExd IIB + H2 (Class 1, Division 1, Group B,C,D)

Electrical Feedthroughs ................................

1/2 in. NPT

Input Power ...................................................

90 to 250 VAC, 50/60 Hz

Power Consumption ......................................

5 VA maximum

External Electrical Noise ...............................

EN 50 082-2, includes 4 kV electrostatic discharge

Handshake Signal to/from

Oxymitter 4000 (self-powered) ................

5 V (5 mA maximum)

Cal Initiate Contact Input from Control Room

5 VDC (self-powered)

Relay Outputs to Control Room ....................

5 to 30 VDC, Form A (SPST)
(one “In-Cal”, one “Cal Failed”)

Cabling Distance between

SPS 4000 and Oxymitter 4000 ................

Maximum 1000 ft (303 m)

Piping Distance between

SPS 4000 and Oxymitter 4000 ................

Maximum 300 ft (91 m)

Approximate Shipping Weight .......................

10 lbs (4.5 kg)

Fisher-Rosemount has satisfied all obligations coming from the European legislation to harmonize
the product requirements in Europe.