Oxymitter 4000 test points -7, Figure 6-6. oxymitter 4000-test points -7, Oxymitter 4000 – Emerson OXYMITTER 4000 User Manual

Page 95: Instruction manual

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0

December 2003

Rosemount Analytical Inc. A Division of Emerson Process Management

Startup and Operation with LOI 6-7

Oxymitter 4000

6-8

OXYMITTER 4000 TEST POINTS

Refer to Figure 6-6. System test points are lo-
cated on the board below the LOI module. Test
points 1 through 6 allow you to monitor with a
multimeter: the heater thermocouple, the O

2

cell

millivolt, and the process O

2

.

a. TP1 and TP2 monitor the oxygen cell

millivolt output which equates to the per-
centage of oxygen present.

b. TP3 and TP4 monitor the heater thermo-

couple.

c. TP5 and TP6 monitor the process gas or

the calibration gas parameter.

6-9

MODEL 751 REMOTE POWERED LOOP
LCD DISPLAY (OPTIONAL)

Refer to Remote Powered Loop LCD manual for
calibration and operation.

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

37260037

Figure 6-6. Oxymitter 4000-Test Points