2-16 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

Oxymitter 4000

-

-

-

-

+

+

+

+

CAL INITIATE

CAL FAIL

LINE IN

IN CAL

NEUTRAL

GROUND

4-20 mA

CONNECTION

NOT USED

NOT USED

5 - 30 VDC TO RELAY OUTPUT

CONNECTIONS

5 VDC

(SELF-POWERED)

TO REMOTE

CONTACT INPUT

CONNECTION

90 - 250 VAC,

50/60 HZ LINE

VOLTAGE

INPUT

FACTORY

WIRING

TO INTERFACE

BOARD

FACTORY WIRING

TO INTERFACE BOARD

FACTORY WIRING

TO POWER SUPPLY

BOARD

FACTORY

WIRING TO

OXYMITTER

4000

FACTORY

WIRING TO

OXYMITTER

4000

26170027

BLACK

WHITE

BROWN

YELLOW

RED

BLUE

ORANGE

GREEN

Figure 2-12. SPS 4000 Electrical Connections

e. Connect

Line

Voltage

Route the line voltage leads into the mani-
fold through the 1/2 in. line voltage conduit
fitting (Figure 2-3) and out through the bot-
tom of the manifold. Connect the LINE IN
and NEUTRAL leads to terminals L and N,
respectively, as shown in Figure 2-12. Also,
be sure to connect the ground wire to the
ground lug. The unit automatically will con-
figure itself for 90 to 250 VAC line voltage
and 50/60 Hz. The power supply requires
no setup.

f. Connect Remote Contact Input Wiring

To set up the SPS 4000 to initiate a calibra-
tion from a remote location, route the 5 VDC
calibration initiate contact input leads
through the 1/2 in. NPT signal conduit port
(Figure 2-3) and out through the bottom of
the manifold. Connect the (+) and (-) CAL
INITIATE leads to terminals 1 and 2, re-
spectively, as shown in Figure 2-12.