Figure 9-5), Figure 9-6), Oxymitter 4000 – Emerson OXYMITTER 4000 User Manual

Page 138: Instruction manual, Figure 9-5. electronic assembly, Tm hart smart family

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0
December 2003

9-10 Maintenance and Service

Rosemount Analytical Inc. A Division of Emerson Process Management

Oxymitter 4000

500 VA

R

SERIAL NO.
TAG NO.

OXYMITTER 4000

VOLTS:

WATTS:

OUTPUT:

LINE FUSE:

Rosemount Analytical Inc.
Orrville, OH 44667-0901

85-264 VAC 48-62 Hz

TM

800-433-6076

4-20 mA

R

5 Amps

TM

HART

SMART FAMILY

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

J1

MOUNTING

SCREW

MOUNTING SCREW

MOUNTING SCREW

26170018

Figure 9-5. Electronic Assembly

3D39619G

REV

1

+

250VAC

TIME LAG

5A

+

+

+

+

+

+

+

+

1

J8

POWER

SUPPLY

BOARD

22220061

Figure 9-6. J8 Connector

7. When reinstalling or replacing the

electronic housing, make sure that O-
ring (10) is in good condition. Place the
J1 and J8 connectors in the hole on the
flat side of the electronic housing.

8. Hold the J1 and J8 connectors out and

to the probe side of the electronic
housing. Make sure that the conduit
port of the electronic housing is on the
same side as the CAL and REF gas

ports. Replace the four screws and
tighten.

9. Reconnect the J8 connector to the

power supply board. Make sure the
connector is secure.

10. Holding the J1 connector leads, slide

the electronic assembly the rest of the
way into the housing. Align the elec-
tronic assembly so that it fits flush on
the pins. To ensure that it is flush,
gently try to rotate the electronics. If
the electronics rotates, repeat the
alignment.

11. Reconnect the J1 connector to the

microprocessor board. Ensure the
connector is secure and tighten the
three captive screws on the microproc-
essor board (top board).

12. Replace the housing cover and ensure

it is tight.

13. Follow the instructions in paragraph

9-4a.2 to install the Oxymitter 4000 into
the stack or duct. If installing an Oxy-
mitter 4000/ SPS 4000 assembly, fol-
low the instructions in paragraph
9-4b.2.