With figure 2-13, E figure 2-14 f, Oxymitter 4000 – Emerson OXYMITTER 4000 User Manual

Page 72: Instruction manual, Figure 2-13. air set, plant air connection, Tm hart smart family

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0
December 2003

2-18 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

Oxymitter 4000

TO ELECTRONICS

REF AIR SET

263C152G01

1

FLOWMETER

0.2-2.0 SCFH 771B635H02

2

2" PRESSURE GAGE

0-15 PSIG

275431-006

3

COMBINATION FILTER-REG.

0-30 PSIG

4505C21G01

NOTE: DIMENSIONS ARE IN INCHES WITH

MILLIMETERS IN PARENTHESES.

1

2

3

4.81 (122.17)

FLOW SET

POINT KNOB

0.125-27 NPT FEMALE

OUTLET CONNECTION

1.19

(30.22)

10.0

(254)

REF

DRAIN VALVE

3.12 (79.25) MAX

8.50

(215.90)

MAX

2.0

(50.80)

2 MOUNTING HOLES

3.19 (81.03) LG

THROUGH BODY FOR

0.312 (7.92) DIA BOLTS

1.50

(38.10)

2.250 (57.15)

SCHEMATIC HOOKUP FOR REFERENCE AIR SUPPLY ON OXYMITTER 4000 PROBE HEAD.

OUTLET

0.25-18 NPT FEMALE

INLET CONNECTION

INSTRUMENT AIR SUPPLY

10-225 PSIG MAX PRESSURE

26170035

0.250 OR 6 MM O.D. TUBING

(SUPPLIED BY CUSTOMER)

500 VA

SERIAL NO.
TAG NO.

OXYMITTER 4000

WATTS:

VOLTS:

FUSE:

LINE

OUTPUT:

Rosemount Analytical Inc.
Orrville, OH 44667-0901

85-264 VAC 48-62 Hz

TM

800-433-6076

4-20 mA

R

5 Amps

TM

HART

SMART FAMILY

Figure 2-13. Air Set, Plant Air Connection

500 VA

R

SERIAL NO.
TAG NO.

OXYMITTER 4000

VOLTS:

WATTS:

OUTPUT:

LINE FUSE:

Rosemount Analytical Inc.
Orrville, OH 44667-0901

85-264 VAC 48-62 Hz

TM

800-433-6076

4-20 mA

R

5 Amps

TM

HART

SMART FAMILY

CALIBRATION GAS

REFERENCE AIR

26170025

Figure 2-14. Oxymitter 4000 Gas Connections Calibration Gas Connections