Instruction Manual

IB-106-340 Rev. 3.0
December 2003

6-6 Setup and Operation with LOI

Rosemount Analytical Inc. A Division of Emerson Process Management

Oxymitter 4000

d. SYSTEM/Status

1. Alarms – Diagnostic alarms. See

Section 8, TROUBLESHOOTING.

2. PID Parameter – Displays the line volt-

age powering the Oxymitter, and infers
the temperature control algorithm being
used to control heater temperature.

3. Reset Device – Device can be reset

here as opposed to repowering. Cali-
bration parameters will be lost.

e. SYSTEM/Software

This is data regarding the Oxymitter 4000
software version, and errors that may have
occurred.

f. SENSOR

DATA

Displays information about the O

2

cell and

thermocouple.

1. Temperatures

(a) O2 Temp – Indicates the thermo-

couple temperature at the sensing
cell this should always be 736

° C.

(b) O2 Temp Max. – Maximum tem-

perature the cell has seen. (Some
process temperatures can exceed
the 736

° C setpoint temperature,

and this will indicate this condition)

(c) Board Temp – The current tem-

perature inside the Oxymitter elec-
tronics housing (85

° C is the max.)

(d) Board Temp Max. – This is the

maximum temperature that the
electronics has experienced over
time.

2. Voltages – The raw MV signals feeding

the temperature indications listed in the
previous paragraph.

3. Output Values – Indication of the cur-

rent readings for O

2

and mA.

6-7

LOI INSTALLATION

The LOI module connects to the top of the elec-
tronic assembly in the electronics housing.
There are four matching connectors (Figure 6-5)
on the back of the LOI module that allow the
user to orient (rotate) the LOI as desired.

37260055

CONNECTOR

RECEPTACLES

LOI MODULE

REAR VIEW

Figure 6-5. LOI Module Connectors