Troubleshooting 8-19

Oxymitter 4000

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

37260033

Alarms

Calib Failed

LOI

KEYPAD

Figure 8-17. Fault 15, Last Calibration Failed

o. Fault 15, Last Calibration Failed

Figure 8-17 shows the electronic assembly
for an Oxymitter 4000 with a membrane
keypad (upper view) and an Oxymitter 4000
with an LOI (lower view).

Membrane Keypad. When Fault 15 is de-
tected, the CALIBRATION LED flashes
three times, pauses for three seconds, and
repeats.

1. The last calibration failed alarm acti-

vates when the slope and constant
values calculated are out of range and
the unit reverts to using the previous
calibration values.

2. The cell should be replaced. See para-

graph 9-4i, Cell Replacement, for cell
replacement instructions.

LOI. When Fault 15 is detected, the LOI
displays the “Calib Failed” message. Refer
to the comments in paragraphs 1 and 2
above.