Startup and Operation with Keypad 5-1

Oxymitter 4000

SECTION 5

STARTUP AND OPERATION

OF OXYMITTER 4000 WITH MEMBRANE KEYPAD

5-1

POWER UP

a. Startup

Display

When power is applied to the probe, the cell
heater turns on. It takes approximately one
half hour for the cell to heat to operating
temperature. This condition is indicated by
the top four LEDs (DIAGNOSTIC ALARMS)
on the membrane keypad (Figure 5-1).
Starting with the CALIBRATION LED, the
LEDs light in ascending order until all four
LEDs are on. At this point, all four turn off
and the cycle starts again. This ramp cycle
continues until the cell is up to operating
temperature.

b. Operating

Display

The ramp cycle turns into a cycle where the
diagnostic LEDs light in sequence from the
top to the bottom, one at a time. After the
bottom LED turns on, the sequence starts
again at the top with the HEATER T/C LED.

c. Error

If there is an error condition at startup, one
of the diagnostics LEDs will be blinking.
Refer Section 7, TROUBLESHOOTING, to
determine the cause of the error. Clear the
error, cycle power, and the operating dis-
play should return.

d. Keypad

The five membrane keys on the membrane
keypad are only used during calibration to
adjust the high and low gas and to initiate
the calibration sequence (Figure 5-2).

5-2

GENERAL OPERATION

a. Overview

Ensure the Oxymitter 4000 is at normal op-
eration. The diagnostic LEDs will display the
operating cycle. All other LEDs should be
off.

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

HEATER T/C

HEATER

O CELL

2

CALIBRATION

LIGHTING SEQUENCE DURING NORMAL OPERATION

(OPERATING DISPLAY)

1

2

3

4

1

2

3

4

HEATER T/C

HEATER

O CELL

2

CALIBRATION

LIGHTING SEQUENCE DURING WARM-UP

(STARTUP DISPLAY)

1

2

3

4

1

2

3

4

22220056

Figure 5-1. Startup and Normal Operation