Oxymitter 4000 setup at the loi -4, Oxymitter 4000, Instruction manual – Emerson OXYMITTER 4000 User Manual

Page 92: And 8% o

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0
December 2003

6-4 Startup and Operation with LOI

Rosemount Analytical Inc. A Division of Emerson Process Management

Oxymitter 4000

SYSTEM

O2 Gas 1

O2 Gas 2

O2-Reset Vals

O2

O2 Cal Intervl

O2-

Gas Time

Purge Time

_____%

_____%

____H

___Sec.

Yes/No

Yes/No

____H

___Sec.

Yes/No

Out Tracks

Next Cal

Auto Calib?

O2 Type

O2 Range

O2 Alarm Leve

_______

______%

l

_____mA

Do O2 Trim

Calib Setup

Input/Output

Parameters

Status

Software

37260018

(CONTINUED FROM

SHEET 1)

Analog

Digital

NOTE

In column four of this menu, the selections in

are user configurable.

text selections are

procedures; related instructions are displayed on the LOI. All other parameters are display only.

Italics

Bold

Version

xxx

Checksum

xxx

Build Number

xxx

Build Date

xxxxxx

Test Code

xx

SW Err File

xx

SW Err Line

xx

Alarms

__________

PID Parameters

115/220

Yes/No

(Cal. required after reset)

Reset Device?

O2 Slope

O2 Constant

O2 T90 Time

Auto Tune?

Lockout Time

Revert Time

Luminance

____ mV/D

____ mV

0:00

Yes/No

0:00

0:00

______

Logic IO Mode

Low

See Table 4-1

O2 Alarm

Force Output

See para. 9-3b

Input State

_______

_______

Figure 6-4. Menu Tree for Local Operator Interface (LOI) on the Oxymitter 4000 (Sheet 2 of 2)

6-6

OXYMITTER 4000 SETUP AT THE LOI

In setting up the Oxymitter 4000 from the LOI,
it’s best to start at the SYSTEM/Calibration
Setup menu, Figure 6-4.

a. SYSTEM/Calibration

Setup

1. O2 Gas #1 – Enter the high or low cal

gas value (the order is not important).

2. O2 Gas #2 – Enter the second cal gas

value.

NOTE

Refer to Section 9, MAINTENANCE AND
SERVICE, for calibration instructions.

NOTE

Rosemount Analytical recommends
0.4% O

2

and 8% O

2

for calibration gases.

3. O2 – Reset Values- resets factory de-

fault values.

4. O2 Output Tracks – 4 to 20 mA signal

can be held at the last value during
calibration, or the signal can be left to
track the cal gases.

5. O2 Cal Interval – If automatic calibra-

tion is selected, this selects the interval
between calibrations.