Figure 8-19 to tr, Figure 8-19, Oxymitter 4000 – Emerson OXYMITTER 4000 User Manual

Page 127: Instruction manual, Symptom — no test gas flow

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0

December 2003

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 8-23

Oxymitter 4000

CHECK ALL WIRING BETWEEN

OXYMITTER 4000 AND SPS 4000.

CHECK FUSE F1 ON POWER

SUPPLY BOARD.

CHECK LOGIC I/O SETTING VIA

HART/AMS.

DISCONNECT CAL GAS INPUT

LINES AT MANIFOLD.

ENSURE CAL GAS FLOWMETER

KNOB IS TURNED COUNTER-

CLOCKWISE TO ALLOW FLOW.

SET LOGIC I/O TO MODE 8 VIA

HART/AMS.

REPLACED CLOGGED CAL GAS

LINE BETWEEN CAL GAS

CYLINDER AND MANIFOLD.

REPLACED CLOGGED CAL GAS

LINE BETWEEN MANIFOLD AND

CAL GAS FLOWMETER.

37260040

PROPERLY CONNECT WIRING OR

SECURE LOOSE WIRING CON-

NECTIONS; REPLACE DAMAGED

WIRING.

REPLACE FUSE PER

PARAGRAPH 9-5a.

SYMPTOM — NO TEST GAS FLOW

NO

NO

NO

NO

YES

YES

YES

YES

YES

NO

CONTINUED

ON SHEET

2 OF 2

IS

WIRING

PROPERLY

CONNECTED

AND

SECURE?

IS

FUSE

BLOWN?

IS

LOGIC I/O

SET FOR

MODE 8?

IS THERE

FLOW?

DOES

CAL GAS

FLOWMETER

REGISTER

FLOW?

POWER SUPPLY BOARD

F1

J4

J3

J2

HI

GAS

LO

GAS

NO

GAS

CAL

RET

J5

Figure 8-19. SPS 4000 Troubleshooting Flowchart (Sheet 1 of 2)