3a t, Recommended configuration -5, Oxymitter 4000 – Emerson OXYMITTER 4000 User Manual

Page 79: Instruction manual

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0

December 2003

Rosemount Analytical Inc. A Division of Emerson Process Management

Configuration with Keypad 3-5

Oxymitter 4000

3-3

RECOMMENDED CONFIGURATION

a. 4-20 mA Signal Upon Critical Alarm

Rosemount recommends that the factory
default be utilized. The 4-20 mA signal will
go to the 3.5 mA level upon any critical
alarm which will cause the O

2

reading to

be unusable. Customer can also select
21.6 mA as the failure setting if normal
operations cause O

2

readings to go below

the zero % O

2

(4 mA) level.

If the O

2

measurement is being utilized as

part of an automatic control loop, the loop
should be placed into manual upon this
failure event or other appropriate action
should be taken.

b. Calibration

Rosemount recommends utilizing an
autocalibration system, actuated by the
“calibration recommended” diagnostic.
New O

2

cells may operate for more than a

year, but older cells may require recali-
bration every few weeks as they near the
end of their life. This strategy ensures
that the O

2

reading is always accurate,

and eliminates many unnecessary cali-
brations based on calendar days or
weeks since previous calibration. When
utilizing the SPS 4000 or the IMPS 4000,
consider wiring some or all associated
alarm contacts.

1. CALIBRATION INITIATE. Contact

from the control room to an SPS
4000 or IMPS 4000 (one per probe)
provides the ability to manually initi-
ate a calibration at any time from the
control room. Note that calibrations

can also be initiated from a HART
handheld communicator, from Asset
Management Solutions software, or
from the keypad on the Oxymitter
4000.

2. IN CALIBRATION. One contact per

probe provides notification to the
control room that the “calibration
recommended” diagnostic has initi-
ated an automatic calibration through
the SPS 4000 or IMPS 4000. If the
O

2

signal is being utilized in an

automatic control loop, this contact
should be utilized to place the control
loop into manual during calibration.

3. CALIBRATION FAILED. One con-

tact per probe from an SPS 4000 or
IMPS 4000 to the control room for
notification that the calibration pro-
cedure failed. Grouped with this
alarm is an output from a pressure
switch which indicates when the cali-
bration gas bottles are empty.

4. 4-20 mA SIGNAL DURING CALI-

BRATION. The 4-20 mA signal can
be configured to respond normally
during any calibration, or can be con-
figured to hold the last O

2

value upon

the initiation of calibration. Factory
default is for the 4-20 mA signal to
operate normally throughout calibra-
tion. Holding the last O

2

value may be

useful if several probes are being av-
eraged for the purpose of automatic
control. Unless several probes are
being averaged, always place any
control loops using the O

2

signal into

manual prior to calibrating.

Advertising