1c to, Gh figure 2-9 pr, Figure 2-8) – Emerson OXYMITTER 4000 User Manual

Page 63: Oxymitter 4000, Instruction manual

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0

December 2003

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-9

Oxymitter 4000

VEE

DEFLECTOR

VEE

DEFLECTOR

DIFFUSION

ELEMENT

SETSCREW

FILTER

GAS FLOW

DIRECTION

APEX

22220020

Figure 2-8. Orienting the Optional Vee Deflector

Uninsulated stacks or ducts may
cause ambient temperatures around
the electronics to exceed 185°F (85°C),
which may cause overheating damage
to the electronics.

8. If insulation is being removed to access

the duct work for Oxymitter 4000
mounting, make sure the insulation is
replaced afterward (Figure 2-9).

c. Remote Electronics Installation

For an Oxymitter 4000 equipped with re-
mote electronics, install the probe according
to the instructions in paragraph 2-1.b. Install
the remote electronics unit on a stand pipe
or similar structure, Figure 2-2.

P

S

U

E

I

T

P

I

C

R

H

W

E

N

T

H

G

C

KE

N

I

-

E

E

R

W

A

V I

S

O

L

P

-

X

O

M

T

A

G

NI

N

-

R

I

T

L

A

I

VE

-

E

E

H

GAS

CAL

.

DRIP
LOOP

LOGIC I/O,
4-20 mA SIGNAL

LINE

VOLTAGE

REPLACE INSULATION

AFTER INSTALLING

OXYMITTER 4000

INSULATION

ADAPTER

PLATE

STACK OR DUCT
METAL WALL

29340005

Figure 2-9. Installation with Drip Loop and Insulation Removal