1e.4 to, Oxymitter 4000, Instruction manual – Emerson OXYMITTER 4000 User Manual

Page 76

Advertising
background image

Instruction Manual

IB-106-340 Rev. 3.0
December 2003

3-2 Configuration with Keypad

Rosemount Analytical Inc. A Division of Emerson Process Management

Oxymitter 4000

c. Verify Oxymitter 4000 Configuration

Located on the microprocessor board, the
top board, are two switches that configure
outputs (Figure 3-2). SW1 determines if the
4-20 mA signal is internally or externally
powered. SW2 determines:

1. Range control, HART or LOCAL.

2. Oxygen range, 0 to 10% O

2

or 0 to

25% O

2

. (0 to 40% O

2

is also configur-

able only through HART/AMS.)

3. The 4-20 mA signal, at fault or power

up, 3.5 mA or 21.6 mA.

Remove power before changing de-
faults. If defaults are changed under
power, damage to the electronics
package may occur.

d. SW1

The two settings are internally or externally
powering the 4-20 mA signal. The factory
setting is for the 4-20 mA signal to be inter-
nally powered.

e. SW2

The factory sets this switch as follows:

1. Position 1 is HART/LOCAL. This switch

controls the configuration of the Oxy-
mitter 4000. The defaults cannot be
changed via HART/AMS unless the
switch is in the HART position. Placing
this switch in the LOCAL position
forces the O

2

range to the setting of

position 2. This switch must be placed
in the LOCAL position or changes in
position 2 will have no effect.

Typically, the probe’s sensing cell,
which is in direct contact with the pro-
cess gases, is heated to approximately
1357°F (736°C), and the external tem-
perature of the probe body may ex-
ceed 842°F (450°C). If operating
conditions also contain high oxygen
levels and combustible gases, the
Oxymitter 4000 may self-ignite.

2. Position 2 determines the O

2

range.

This can be set to either 0 to 10% O

2

or 0 to 25% O

2

. The factory setting is 0

to 10% O

2

. If necessary, the O

2

range

can be configured from 0 to 40% O

2

.

To select values within this range, set
position 1 of SW2 to HART and then
enter the range via HART/AMS. Do not
change position 1 of SW2 to LOCAL
unless you want to operate in the
range specified by position 2 of SW2.

3. Position 3 determines the output at

startup or at an alarm. The settings are
3.5 mA or 21.6 mA. The factory setting
is 3.5 mA. At startup, the current at the
analog output is 3.5 mA or 21.6 mA.

4. Position 4 can be used to set the

heater for 115 or 220 VAC operation.
This switch is functional only when the
software is set for manual voltage se-
lection (Auto Tune = No). Otherwise,
the internal electronics auto detect the
input line voltage and sets the heater
voltage accordingly (Auto Tune = Yes).

f. Once the cell is up to operating temperature,

the O

2

percentage can be read:

1. Access TP5 and TP6 next to the mem-

brane keypad. Attach a multimeter
across TP5 and TP6. The calibration
and process gases can now be moni-
tored. Pressing the INC or DEC once
will cause the output to switch from the
process gas to the calibration gas.
Pressing INC or DEC a second time
will increase or decrease the calibra-
tion gas parameter. If the keys have
been inactive for one minute, the out-
put reverts to the process gas. When a
calibration has been initiated, the value
at TP5 and TP6 is the % O

2

seen by

the cell. Oxygen levels, as seen on the
multimeter, are:

8.0% O2 = 8.0 VDC
0.4% O2 = 0.4 VDC

2. HART/AMS.

3. Model 751. The loop-driven LCD

display.

Advertising