Installation 2-7

Oxymitter 4000

NOTE: ALL MASONRY STACK WORK AND JOINTS EXCEPT

ADAPTER PLATE NOT FURNISHED BY ROSEMOUNT.

INSTALLATION FOR MASONRY

WALL STACK CONSTRUCTION

INSTALLATION FOR METAL

WALL STACK OR DUCT

CONSTRUCTION

0.50 [13]

3.75 [95]

MIN DIA HOLE

IN WALL

STACK OR DUCT

METAL WALL

MTG HOLES

SHOWN ROTATED

45 OUT OF

TRUE POSITION

o

WELD OR BOLT ADAPTER

PLATE TO METAL WALL

OF STACK OR DUCT.

JOINT MUST BE AIRTIGHT.

0.50 [13]

4.50 [114]

O.D. REF

PIPE 4.00 SCHED 40

PIPE SLEEVE (NOT

BY ROSEMOUNT)

LENGTH BY CUSTOMER

MASONRY

STACK WALL

OUTSIDE WALL

SURFACE

JOINT MUST

BE AIRTIGHT

MTG HOLES

SHOWN ROTATED

45 OUT OF

TRUE POSITION

o

FIELD WELD

PIPE TO

ADAPTER PLATE

BOLT ADAPTER

PLATE TO OUTSIDE

WALL SURFACE

NOTE: DIMENSIONS IN INCHES WITH

MILLIMETERS IN PARENTHESES.

2.50 [63.5]

MIN DIA HOLE

IN WALL

STACK OR DUCT

METAL WALL

WELD OR BOLT ADAPTER

PLATE TO METAL WALL

OF STACK OR DUCT.

JOINT MUST BE AIRTIGHT.

FIELD WELD

PIPE TO

ADAPTER PLATE

3.50 [89]

O.D. REF

PIPE 3.00 SCHED 40

PIPE SLEEVE (NOT

BY ROSEMOUNT)

LENGTH BY CUSTOMER

MASONRY

STACK WALL

OUTSIDE WALL

SURFACE

JOINT MUST

BE AIRTIGHT

BOLT ADAPTER

PLATE TO OUTSIDE

WALL SURFACE

35830005

Figure 2-6. Oxymitter 4000 Adapter Plate Installation