5 what to do if... / printer, What to do if... / printer – YSI TruLab 1320P User Manual

Page 16

Advertising
background image

Printer (only TruLab pH/ISE 1320 P)

TruLab pH/ISE 1320

16

ba76157e01

03/2014

4.5

What to do if... / printer

Integrated printer

does not print

Printer operating -

paper not being

printed

Integrated printer

prints automatically

Cause

Remedy

– Printer switched off

(LED flashing)

– Switch on the printer

(LED illuminated)

– No power pack connected

– Connect the power pack

– USB cable connected

– Disconnect the USB cable from the

meter

– Function "store automatically at

intervals" with long intervals is
switched on

– Switch off the function (see section

10.2 A

UTOMATIC

DATA

STORAGE

AT

INTERVALS

, page 79)

– No paper available

– Insert a roll of paper

Cause

Remedy

– Paper inserted the wrong way up – Turn the roll of paper around and

insert it with the other side upwards

Cause

Remedy

– The function "store automatically

at intervals" or "transmit data
automatically at intervals" is
switched on

– Switch off the functions (see sec-

tion 10.2 A

UTOMATIC

DATA

STORAGE

AT

INTERVALS

, page 79 or

section 11.1 O

PTIONS

FOR

DATA

TRANSMISSION

, page 84)

Advertising