TruLab pH/ISE 1320

Settings

ba76157e01

03/2014

67

Calibration/

Serial number (sen-
sor)

-

Entry of the series number of the
connected sensor.
The series number is output in the
calibration record.

Change the contents of the high-
lighted position with <><>.

Go to the next digit with <F2>/
[].

After theseries number was
completely entered, confirm with
<ENTER>.

Calibration /Buffer

YSI
ConCal
NIST/DIN
...

Buffer sets to be used for pH cali-
bration (see section 6.2 C

ALIBRA

-

TION

, page 31).

Calibration /One
point calibration

Yes
No

Quick calibration with 1 buffer

Calibration /Cali-
bration interval

1 ... 7 ... 999 d

Calibration interval for the pH elec-
trode (in days).
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring screen.

Calibration /Unit for
slope

mV/pH
%

Unit of the slope.
The % display refers to the Nernst
slope of -59.2 mV/pH (100 x deter-
mined slope/Nernst slope).

Man. temperature

-25 ... +25 ...
+130 °C

Entry of the manually determined
temperature. For measurements
without temperature sensor only.

Temperature unit

°C
°F

Temperature unit,

degrees Celsius or degrees Fahr-
enheit.
All temperature values are dis-
played with the selected unit.

Resolution pH

0.001
0.01
0.1

Resolution of the pH display