9 settings, 1 measurement settings, 1 settings for ph measurements – YSI TruLab 1320P User Manual

Page 66: Settings, Measurement settings, Settings for ph measurements, Selected (see section 9.1.1 s, Ettings, Measurements, Section 9.1.1 s

Advertising
background image

Settings

TruLab pH/ISE 1320

66

ba76157e01

03/2014

9

Settings

9.1

Measurement settings

9.1.1 Settings for pH measurements

The settings for pH measurements are made in the menu for calibration and
measurement settings of the pH/ORP measurement. To open the settings, dis-
play the required measured parameter in the measured value display and
press the <MENU/ENTER> key. After completing the settings, switch to the
measured value display with <M>.
In the following table, only those settings are listed that concern the pH mea-
surement.
Default settings are printed in bold.

Menu item

Possible set-
ting

Description

Calibration /Cali-
bration record

-

Displays the calibration record of
the last calibration.

Calibration /
Calibration data
storage
/Display

-

Displays the calibration record.

Further options:

Scroll through the calibration
records with <><>.

Output the displayed calibration
record to the interface with
<F2>/[USB output].

Output all calibration records to
the interface with <F2__>[USB
output]
.

Quit the display with <F1>/
[Back] or <MENU/ENTER>.

Switch directly to the measured
value display with <M>.

Calibration /

Calibration data
storage
/

Output to USB

-

Outputs the calibration records to
the interface.