What to do if...

TruLab pH/ISE 1320

90

ba76157e01

03/2014

No stable measured

value

Obviously incorrect

measured values

Cause

Remedy

pH electrode:

– Junction contaminated

– Clean the junction

– Membrane contaminated

– Clean the membrane

Test sample:

– pH value not stable

– Measure with air excluded if

necessary

– Temperature not stable

– Temper if necessary

Electrode + test sample:

– Conductivity too low

– Use a suitable electrode

– Temperature too high

– Use a suitable electrode

– Organic liquids

– Use a suitable electrode

Cause

Remedy

pH electrode:

– pH electrode unsuitable

– Use a suitable electrode

– Temperature difference between

buffer and test sample too great

– Adjust the temperature of buf-

fer or sample solutions

– Measurement procedure not suit-

able

– Follow a special procedure