YSI TruLab 1320P User Manual

Page 52

Advertising
background image

Ion concentration

TruLab pH/ISE 1320

52

ba76157e01

03/2014

Calibration

evaluation

After calibrating, the meter automatically evaluates the calibration.

Calibration record

(example)

Menu item

Setting/func-
tion

Description

Calibration /
Calibration data stor-
age
/Display

-

Displays the calibration record.

Further options:

Scroll through the calibration
records with <><>.

Output the displayed calibra-
tion record to the interface with
<F2>/[USB output].

Quit the display with <F1>/
[Back] or <MENU/ENTER>.

Switch directly to the mea-
sured value display with <M>.

Calibration /

Calibration data stor-
age
/

Output to USB

-

Outputs the calibration records to
the interface.

Display

Calibration
record

Magnitude of the slope [mV]

+++

50.0 ... 70.0 or 25.0 ... 35.0

Error

Error

< 50 or > 70
or
< 25 or > 35

Error elimination (see section 13 W

HAT

TO

DO

IF

..., page 89)

TruLab pH/ISE 1320
Ser. no. 12345678

CALIBRATION ISE
18.01.2013 08:09:10

Sensor
Ser. no. 12345678

Standard 1

0.010 mg/l

Standard 2

0.020 mg/l

Voltage 1

38.5 mV

Voltage 2

58.0 mV

Temperature 1

24.0 øC

Temperature 2

24.0 øC

Ion type

Ag

Slope

64.7 mV

Sensor

+++

_____________________________________

Advertising
This manual is related to the following products: