pH value

TruLab pH/ISE 1320

36

ba76157e01

03/2014

Continuing with

three- to five-point

calibration

14.

Immerse the electrode in the second buffer solution.

15.

When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.

16.

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.

17.

Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>.

The calibration display for the setting of the nominal buffer value
appears.

18.

Set the nominal buffer value for the measured temperature with
<><>.

19.

Accept the set calibration value with <ENTER>.

The calibration display for the next buffer appears (voltage display).

20.

Finish the calibration procedure as a two-point calibration with <M>.
The calibration record is displayed.

pH

4.035

24.8 °C

ConCal

2

12.12.2012

12:12

Buffer



21.

Thoroughly rinse the electrode with deionized water.

22.

Immerse the electrode in the next buffer solution.

23.

When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.

24.

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.

25.

Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>.

The calibration display for the setting of the nominal buffer value
appears.