YSI TruLab 1320P User Manual

Page 64

Advertising
background image

Ion concentration

TruLab pH/ISE 1320

64

ba76157e01

03/2014

Measuring

CALIBRATION
12.12.2012 12:12:12
#1 0.010 mg/l

20.2 mV 25.0 °C

#2 0.100 mg/l

79.2 mV 25.0 °C

Slp.: 59.0 mV
Sensor +++ (Ion type: Ag)

12.12.2012

12:12

Continue

USB output

5.

Start the measurement with <F1> Continue

An entry window appears.

6.

Prepare the sample.

7.

If necessary, add the ISA/TISAB solution to the sample.

8.

Thoroughly rinse the electrode with deionized water.

9.

Immerse the electrode in the sample that was supplemented with blank
value solution.

10.

Using <> <> and <MENU/ENTER>, enter the values for the vol-
ume of the sample (Sample volume), the volume of the ISA/TISAB
solution (ISA/TISAB vol.), the volume of the blank value solution (BV
volume
) and the concentration of the blank value solution (BV conc.).

11.

Select Continue with <> <> and start the measurement with
<MENU/ENTER>.

When the measurement is finished an entry window appears.

i

Immerse sensor in sample

Sample volume

100.0 ml

ISA/TISAB vol.

1.0 ml

BV volume

100.0 ml

BV conc.

1.000 mg/l

Continue

Blank value addition

12.12.2012

12:12

Back

i

Add standard!

Std. volume

100.0 ml

Std. conc.

1.0 mg/l

Continue

Blank value addition

12.12.2012

12:12

Back