Ion concentration

TruLab pH/ISE 1320

62

ba76157e01

03/2014

Measuring

7.

Start the measurement with <F1> Continue.

An entry window appears.

8.

Prepare the standard solution.

9.

If necessary, add the ISA/TISAB solution to the standard solution.

10.

Thoroughly rinse the electrode with deionized water.

11.

Immerse the electrode in the standard solution (complexing agent or
precipitating agent).

12.

Using <> <> and <MENU/ENTER>, enter the values for the vol-
ume of the standard solution (Std. volume), the concentration of the
standard solution (Std. conc.) and the volume of the ISA/TISAB solu-
tion (ISA/TISAB vol.).

13.

Select Continue with <> <> and start the measurement with
<MENU/ENTER>.

When the measurement is finished an entry window appears.

14.

Add the sample to the standard (complexing agent or precipitating
agent).

15.

Using <> <> and <MENU/ENTER>, enter the value for the volume
of the sample (Sample volume).

16.

Select Continue with <> <> and start the measurement with
<MENU/ENTER>.

The measurement result is displayed when the measurement is com-
pleted.

i

Immerse sensor in standard

Std. volume

100.0 ml

Std. conc.

1.0 mg/l

ISA/TISAB vol.

1.0 ml

Continue

Sample subtraction

12.12.2012

12:12

Back

i

Add sample!

Sample volume

1.0 ml

Continue

Sample subtraction

12.12.2012

12:12

Back