YSI TruLab 1320P User Manual

Page 55

Advertising
background image

TruLab pH/ISE 1320

Ion concentration

ba76157e01

03/2014

55

11.

Select Continue with <> <> and start the measurement with
<MENU/ENTER>.

When the measurement is finished an entry window appears.

12.

Add the standard solution to the sample.

13.

Using <> <> and <MENU/ENTER>, enter the values for the vol-
ume of the standard solution (Std. volume) and concentration of the
standard solution (Std. conc.) .

14.

Select Continue with <> <> and start the measurement with
<MENU/ENTER>.

The measurement result is displayed when the measurement is com-
pleted.

15.

If necessary, start measuring further samples with <MENU/ENTER>.

Repeat steps 6 - 14 for all samples.

16.

Terminate the measuring method with <M>.

A safety query appears.

17.

Select Yes with <> <>.

18.

Confirm Yes with <MENU/ENTER>.

Measurement with the selected method is completed.

i

Add standard!

Std. volume

1.0 ml

Std. conc.

1.0 mg/l

Continue

Standard addition

12.12.2012

12:12

Back

ISE

0.321

24.8 °C

Ag

12.12.2012

12:12

Standard addition

mg/l

USB output

Advertising