TruLab pH/ISE 1320

Ion concentration

ba76157e01

03/2014

57

10.

Using <> <> and <MENU/ENTER>, select the values for the vol-
ume of the sample (Sample volume) and the volume of the ISA/TISAB
solution (ISA/TISAB vol.).

11.

Select Continue with <> <> and start the measurement with
<MENU/ENTER>.

When the measurement is finished an entry window appears.

12.

Add the standard (complexing agent or precipitating agent) to the sam-
ple.

13.

Using <> <> and <MENU/ENTER>, enter the values for the vol-
ume of the standard solution (Std. volume) and concentration of the
standard solution (Std. conc.) .

14.

If necessary, set the ion type for the ion in the standard (Ion type) with
<> <> and <MENU/ENTER>.

On selection of the ion type that can be defined (ION):

Using <> <> and <MENU/ENTER>, set the valence (Valency) and
molar weight (Molar mass) for the ion in the standard solution.

15.

Select Continue with <> <> and start the measurement with
<MENU/ENTER>.

The measurement result is displayed when the measurement is com-
pleted.

16.

If necessary, start measuring further samples with <MENU/ENTER>.

Repeat steps 6 - 13 for all samples.

17.

Terminate the measuring method with <M>.

A safety query appears.

18.

Select Yes with <> <>.

i

Add standard!

Ion type

S

Std. volume

1.0 ml

Std. conc.

1.0 mg/l

Continue

Standard subtraction

12.12.2012

12:12

Back

ISE

0.321

24.8 °C

Ag

12.12.2012

12:12

Standard subtraction

mg/l

USB output