YSI TruLab 1320P User Manual

Page 35

Advertising
background image

TruLab pH/ISE 1320

pH value

ba76157e01

03/2014

35

Continuing with two-

point calibration

7.

When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.

8.

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.

9.

Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.

The calibration display for the setting of the nominal buffer value
appears.

10.

Set the nominal buffer value for the measured temperature with
<><>.

11.

Accept the set calibration value with <ENTER>.

The calibration display for the next buffer appears (voltage display).

12.

If necessary, finish the calibration procedure as a single-point calibration
with <M>.
The calibration record is displayed.

For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the elec-
trode.

pH

0.0

24.8 °C

ConCal

7.000

12.12.2012

12:12

Buffer

mV

AR

pH

7.000

24.8 °C

ConCal

1

12.12.2012

12:12

Buffer



13.

Thoroughly rinse the electrode with deionized water.