6 ph value, 1 measuring, 1 measuring the ph value – YSI TruLab 1320P User Manual

Page 28: Ph value, Measuring, Measuring the ph value, 6ph value

Advertising
background image

pH value

TruLab pH/ISE 1320

28

ba76157e01

03/2014

6

pH value

6.1

Measuring

6.1.1 Measuring the pH value

NOTE

When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as the values would be incorrect
.

The USB interface is not galvanically isolated.

Stability control

(AutoRead)

& HOLD function

The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values.
The measured parameter flashes on the display

as soon as the measured value is outside the stability range

when the automatic Stability control is switched off.

You can activate or switch off the automatic Stability control function (see
section 9.3.1 S

YSTEM

, page 75).

To ensure the high measurement accuracy of the measuring sys-
tem, always measure with a calibrated electrode (see section 6.2
C

ALIBRATION

, page 31).

1.

Connect the pH electrode to the meter.

2.

If necessary, select the measured parameter with <M>.

3.

When measuring without temperature sensor:
Temper the test sample or measure the current temperature and enter it
with <> <>.

4.

Immerse the pH electrode in the test sample.
The measured value is checked for stability (automatic stability control).
The display of the measured parameter flashes.

5.

Wait for a stable measured value.
The display of the measured parameter no longer flashes.

pH

6.949

24.8 °C

12.12.2012

12:12

USB output

Advertising
This manual is related to the following products: