YSI TruLab 1320P User Manual

Page 50

Advertising
background image

Ion concentration

TruLab pH/ISE 1320

50

ba76157e01

03/2014

Continuing with

three- to seven-point

calibration

Repeat the steps 8 to 13 in the same way with the third and further standard
solutions as necessary. The new calibration values are displayed after the last
calibration step was completed.

14.

When calibrating without temperature sensor:

Measure the temperature of the standard solution using a thermom-
eter.

Use <F2>/[ T

⇅ C ] to select the setting of the temperature.

Use <> <> to set the temperature.

Use <F2>/[ T

⇅ C ] to select the setting of the concentration.

15.

Set the concentration of the standard solution with <> <> and
press <MENU/ENTER>.

The standard solution is measured.

The measured value is checked for stability (AutoRead).

16.

Wait for the end of the AutoRead measurement or accept the calibra-
tion value with <MENU/ENTER>.

The calibration display for the next standard solution appears.

17.

Press <MENU/ENTER> to continue with three-point calibration.
or
Finish the calibration procedure as a two-point calibration with <M>.

The new calibration values are displayed.

ISE

8.9

24.8 °C

12.12.2012

12:12

0.01 m

mV

AR

ISE

20.5

24.8 °C

Ag

0.050

mg/l

3

12.12.2012

12:12

Standard

mV

Based on the calibration data, the calibration curve is determined
in sections, according to the Nernst equation modified by Nikolski.

Advertising
This manual is related to the following products: