6 calibration data, Calibration data – YSI TruLab 1320P User Manual

Page 38

Advertising
background image

pH value

TruLab pH/ISE 1320

38

ba76157e01

03/2014

6.2.6 Calibration data

Displaying the

calibration data

The calibration record of the last calibration is to be found under the menu item,
<MENU/ENTER> / Calibration / / Calibration record. To open it, press the
<CAL__> key in the measured value display.

Subsequently, you can transmit the displayed calibration data to the interface,
e. g. to a PC, with <F2>[USB output].

Displaying the

calibration data

storage

The calibration records of the last calibrations (up to 10) are given in the menu,
<MENU/ENTER>/Calibration / Calibration data storage.

Calibration

evaluation

After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.

The calibration record is automatically transmitted to the interface
after calibrating.

Menu item

Setting/func-
tion

Description

Calibration /
Calibration data stor-
age
/Display

-

Displays the calibration record.

Further options:

Scroll through the calibration
records with <><>.

Output the displayed calibra-
tion record to the interface with
<F2>/[USB output].

Quit the display with <F1>/
[Back] or <MENU/ENTER>.

Switch directly to the mea-
sured value display with <M>.

Calibration /

Calibration data stor-
age
/

Output to USB

-

Outputs the calibration records to
the interface.