TruLab pH/ISE 1320

pH value

ba76157e01

03/2014

39

Calibration record

(example)

Display

Calibration
record

Zero point
[mV]

Slope

[mV/pH]

+++

- 15 ... + 15

-60.5 ... -58.0

++

- 20 ... <-15
or
>+15 ... + 20

>-58.0 ... -57.0

+

- 25 ... <- 20
or
>+ 20 ... + 25

-61.0 ... <-60.5
or
>-57.0 ... -56.0

-

- 30 ... <- 25
or
->+25 ... + 30

-62.0 ... <-61.0
or
>-56.0 ... -50.0

Clean the electrode according to the elec-
trode operating manual

Error

Error

- 30 ... + 30

-62.0 ... -50.0

Error elimination (see section 13 W

HAT

TO

DO

IF

..., page 89)

TruLab pH/ISE 1320
Ser. no. 11292113

CALIBRATIONpH
12.12.2012 15:55

Ser. no. 10501234
AutoCal TEC
Buffer 1

4.01

Buffer 2

7.00

Buffer 3

10.01

Voltage 1

184.0 mV

Voltage 2

3.0 mV

Voltage 3

-177.0 mV

Temperature 1

24.0 °C

Temperature 2

24.0 °C

Temperature 3

24.0 °C

Slope

-60.2 mV/pH

Asymmetry

4.0 mV