3 calibration interval, Calibration interval – YSI TruLab 1320P User Manual

Page 71

Advertising
background image

TruLab pH/ISE 1320

Settings

ba76157e01

03/2014

71

* Brand names or trade names are trademarks of their respective owners protected

by law.

9.1.3 Calibration interval

The calibration evaluation is displayed as a sensor symbol.

The sensor symbol flashes after the adjusted calibration interval has expired. It
is still possible to measure.

Setting the

calibration interval

The calibration interval is set to 7 days in the factory.

You can change the interval (1 ... 999 days):

23

Reagecon 25 *

2.000
4.000
7.000
10.000
13.000

25 °C

24

Chemsolute *

2.000
4.000
7.000
10.000

20°C

25

USABlueBook

*

4.000
7.000
10.000

25 °C

No.

Buffer set *

pH values

at

The buffers are selected in the menu, pH / <MENU/ENTER> / Cali-
bration
/ Buffer (see 9.1.1 S

ETTINGS

FOR

P

H

MEASUREMENTS

,

PAGE

66).

To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.

1.

Open the menu for measurement settings with <MENU/ENTER>.

2.

In the Calibration / Calibration interval menu, set the calibration interval
with <><>.

3.

Confirm the setting with <ENTER>.

4.

Quit the menu with <M>.