YSI TruLab 1320P User Manual

Page 33

Advertising
background image

TruLab pH/ISE 1320

pH value

ba76157e01

03/2014

33

Continuing with

three- to five-point

calibration

14.

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.

15.

Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>.

The calibration display for the next buffer appears (voltage display).

16.

If necessary, finish the calibration procedure as a two-point calibration
with <M>.
The calibration record is displayed.

pH

0.0

24.8 °C

AutoCal TEC

7.000

12.12.2012

12:12

AR

Buffer

mV

17.

Thoroughly rinse the electrode with deionized water.

18.

Immerse the electrode in the next buffer solution.

19.

When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.

20.

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.

pH

180.0

24.8 °C

AutoCal TEC

4.010

12.12.2012

12:12

AR

Buffer

mV