7 continuous measurement control (cmc function), Continuous measurement control (cmc function) – YSI TruLab 1320P User Manual

Page 40

Advertising
background image

pH value

TruLab pH/ISE 1320

40

ba76157e01

03/2014

6.2.7 Continuous measurement control (CMC function)

The Continuous Measurement Control (CMC function) facilitates to evaluate
the current measured value instantly and definitely.

After each successful calibration the scale of the pH measuring range is
displayed in the measured value display. Here you can very clearly see
whether or not the current measured value is in the calibrated part of the
measuring range.

The following information is displayed:

The limits of the calibrated range are determined by the buffers used for cali-
bration:

1 Measuring range for which a valid calibration is available (white).

Measured values in this range are suitable for documentation.

2 Measuring range for which no valid calibration is available (shaded).

Measured values in this range are not suitable for documentation. If
necessary, calibrate the meter with buffers covering this measuring
range.

If the current measured value is outside the calibrated range, this
area is shaded stronger.

If a measured value is outside the measuring range pH 0 - 14, over-
flow arrows are displayed at the left or right edge of the measuring
range.

3 Currently measured pH value (needle)
4 Marking lines for all nominal buffer values used with the last valid cali-

bration

Lower limit:

Buffer with lowest pH value - 2 pH units

Upper limit:

Buffer with highest pH value + 2 pH units

25.0

6.093

1

2

3

4

pH

°C

0

14

12.12.2012

12:12

USB output

Advertising
This manual is related to the following products: