2 calibration, 1 why calibrate, 2 when to calibrate – YSI TruLab 1320P User Manual

Page 48: 3 calibration (ise cal), Calibration, Why calibrate, When to calibrate, Calibration (ise cal)

Advertising
background image

Ion concentration

TruLab pH/ISE 1320

48

ba76157e01

03/2014

8.2

Calibration

8.2.1 Why calibrate?

Ion-selective electrodes age and are temperature-dependent. This changes
the slope. As a result, an inexact measured value is displayed. Calibration
determines the calibration line of the electrode and stores this value in the
meter.

Thus, you should calibrate before each measurement and at regular intervals.

8.2.2 When to calibrate?

Before any measurement if possible

After connecting another ISE electrode

When the sensor symbol flashes, e.g. after a voltage interruption (empty
batteries)

8.2.3 Calibration (ISE Cal)

ISE Cal is the conventional two-point to seven-point calibration procedure
that uses 2 to 7 freely selectable standard solutions. The concentration
expected for the measurement determines the concentration of the calibration
standards.

1.

Connect the ISE combination electrode to the meter.
The pH/mV/ISE measuring window is displayed.

2.

Keep the standard solutions ready.

3.

If necessary, measure the temperature of the standard solutions with a
thermometer.

4.

If necessary, change the unit of the measurement result and calibration
standards in the ISE setup/Unit menu.

5.

In the measured value display, select the ISE measuring window with
<> <> and <M>.

6.

Start the calibration with <CAL>.

The calibration display appears.

ISE

0.0

24.8 °C

Ag

0.010

mg/l

1

12.12.2012

12:12

Standard

mV

Advertising
This manual is related to the following products: