4 calibration standards, 5 calibration data, Calibration standards – YSI TruLab 1320P User Manual

Page 51: Calibration data

Advertising
background image

TruLab pH/ISE 1320

Ion concentration

ba76157e01

03/2014

51

8.2.4 Calibration standards

Use two to seven different standard solutions. The standard solutions have to
be selected in either increasing or decreasing order.

8.2.5 Calibration data

Displaying the

calibration data

The calibration record of the last calibration is to be found under the menu item,
<MENU/ENTER> / Calibration / / Calibration record. To open it, press the
<CAL__> key in the measured value display.

Subsequently, you can transmit the displayed calibration data to the interface,
e. g. to a PC, with <F2>[USB output].

Displaying the

calibration data

storage

The calibration records of the last calibrations (up to 10) are given in the menu,
<MENU/ENTER>/ ISE / Calibration / Calibration data storage.

Select the unit of the standard solution and measurement result in
the ISE setup/Unit menu.

Standard solution (Std 1 - 7) Values

Unit [mg/l]

0.010 ... 500,000

Unit [mol/l]

0.100 ... 5,000 µmol/l
10.00 ... 5,000 mmol/l

Unit [mg/kg]

0.010 ... 500,000

Unit [ppm]

0.010 ... 500,000

Unit [%]

0.001 ... 50,000

The measurement precision is also dependent on the selected
standard solutions. Therefore, the selected standard solutions
should cover the value range expected of the subsequent concen-
tration measurement.

If the measured electrode potential is outside the calibrated range,
the [ISEErr] warning is displayed.

Advertising
This manual is related to the following products: