2 standard subtraction, Stan – YSI TruLab 1320P User Manual

Page 56

Advertising
background image

Ion concentration

TruLab pH/ISE 1320

56

ba76157e01

03/2014

8.3.2 Standard subtraction

In the "Standard Subtraction" procedure, a known amount of standard solution
is added to the sample (as complexing agent or precipitating agent) and, thus,
the ion concentration lowered.
The ion concentration in the sample is calculated from the change in potential.

Calibration

Measuring

1.

Select the measuring method (see section 8.3 S

ELECTING

THE

MEASUR

-

ING

METHOD

, page 53).

2.

Prepare two standard calibration solutions.

3.

Perform a two-point calibration according to user guidance.

4.

The calibration record is displayed as soon as a stable value is
achieved for the second standard calibration solution.

CALIBRATION
12.12.2012 12:12:12
#1 0.010 mg/l

20.2 mV 25.0 °C

#2 0.100 mg/l

79.2 mV 25.0 °C

Slp.: 59.0 mV
Sensor +++ (Ion type: Ag)

12.12.2012

12:12

Continue

USB output

5.

Start the measurement with <F1>/[Continue].

An entry window appears.

6.

Prepare the sample.

7.

If necessary, add the ISA/TISAB solution to the sample.

8.

Thoroughly rinse the electrode with deionized water.

9.

Immerse the electrode in the sample.

i

Immerse sensor in sample

Sample volume

100.0 ml

ISA/TISAB vol.

1.0 ml

Continue

Standard subtraction

12.12.2012

12:12

Back