2 resetting the system settings, Resetting the system settings – YSI TruLab 1320P User Manual
Page 77
TruLab pH/ISE 1320
Settings
ba76157e01
03/2014
77
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open it in the measured value display,
press the <MENU/ENTER> key.
9.4.2 Resetting the system settings
The following system settings can be reset to the delivery status:
The system settings are reset in the menu, Storage & config / System / Reset.
To open the Storage & config menu, press the <MENU/ENTER__> key in the
measured value display.
Resolution pH
0.001
Resolution mV
0.1
Asymmetry
0 mV
Slope
-59.2 mV
Man. temperature
25 °C
One point calibration
No
Stability control
On
Temperature unit
°C
Setting
Default settings
Language
English
Beep
On
Baud rate
4800 Baud
Output format
ASCII
Decimal separator
Dot (xx.x)
Contrast
50 %
Illumination
Auto
Switchoff time
1 h
Setting
Default settings