TruLab pH/ISE 1320

Data storage

ba76157e01

03/2014

83

Quitting the display

To quit the display of stored measurement datasets, you have the following
options:

Switch directly to the measured value display with <M>.

Quit the display and move to the next higher menu level with <F1>/[Back].

10.3.2 Erasing the measurement data storage

Erasing the measurement data storage (see section 10.3.1 E

DITING

THE

MEASUREMENT

DATA

STORAGE

, page 81).

10.3.3 Measurement dataset

A complete dataset includes:

ID number

Date/time

Measured values of the connected sensors

Measured temperature value of the connected sensors or manually set tem-
perature

AutoRead info: The AR indicator appears with the measured value if the
AutoRead criterion was met while storing (stable measured value). Other-
wise, there is no AR indicator.

Calibration evaluation: +++, ++, +, -, or no evaluation

10.3.4 Storage locations

The TruLab pH/ISE 1320 meter has two measurement data storages. The
measured values recorded either manually or automatic are stored separately
in individual measurement data storages.

Storage

Maximum number of datasets

Manual data storage

200

Automatic data storage

5000