2 connecting a pc, 3 multilab importer, 2 connecting a pc 11.3 multilab importer – YSI TruLab 1320P User Manual

Page 85

Advertising
background image

TruLab pH/ISE 1320

Transmitting data (USB interface)

ba76157e01

03/2014

85

11.2 Connecting a PC

Connect the TruLab pH/ISE 1320 to the PC via the USB interface.

NOTE

When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as the values would be incorrect
.

The USB interface is not galvanically isolated.

Installation of the

USB driver on the

PC

System requirements of the PC for installation of the USB driver:

PC with Pentium processor or higher with at least one free USB connection
and CD-ROM drive

Windows XP, Windows 7.

11.3 MultiLab Importer

With the aid of the MultiLab Importer software, you can record and evaluate
measurement data with a PC.

1.

Insert the supplied installation CD in the CD drive of your PC.

2.

Install the driver from the CD.
Follow the Windows installation instructions as necessary.

3.

Connect the TruLab pH/ISE 1320 to the PC via the USB interface.
The meter is listed as a virtual COM interface among the connections in
the Windows instrument manager.

More detailed information can be found in the MultiLab Importer
operating manual.