TruLab pH/ISE 1320

What to do if...

ba76157e01

03/2014

91

13.2 ISE

Error message OFL

Obviously incorrect

measured values

Error message,

Error (invalid

calibration)

Warning [TpErr]

Warning [ISEErr]

Cause

Remedy

– Measuring range exceeded

– Dilute the test sample

Cause

Remedy

– Electrode not connected

– Connect the electrode

– Cable broken

– Replace the electrode

Cause

Remedy

ISE electrode:

– Moisture in the plug

– Dry the plug

– Electrode obsolete

– Replace the electrode

– Electrode not suitable for the range

to be measured

– Use a suitable electrode

– Electrode not suitable for the

selected ion

– Use a suitable electrode or

select a suitable ion

– The gas-sensitive electrode NH 500

was calibrated with the NH4 setting

– Select the following settings:

Ion type "ION", Valency "-1"

– Socket damp

– Dry the socket

Calibration procedure:

– Wrong sequence of standards for

three point calibration

– Select the correct sequence

– Calibration standards do not have

the correct temperature (max.

±

2 °C

temperature difference)

– Adjust the temperature of the

calibration standards

Cause

Remedy

– Temperature difference between

measurement and calibration
greater than 2 K.

– Adjust the temperature of the

test sample

Cause

Remedy

– Electrode voltage outside calibrated

range

– Recalibrate