TruLab pH/ISE 1320

What to do if...

ba76157e01

03/2014

93

13.3 General information

Sensor symbol

flashes

display

Meter does not react

to keystroke

You want to know

which software

version is in the

meter

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– Batteries almost empty

– Replace the batteries (see

section 12.1 M

AINTENANCE

,

page 87)

Cause

Remedy

– Operating condition undefined or

EMC load unallowed

– Processor reset:

Press the <ENTER> and <On/
Off>
key simultaneously

Cause

Remedy

– E. g., a question by the service

department

– Switch on the meter.

Open the menu, <ENTER_> /
Storage & config / System /
Service information. The
instrument data are displayed.