2 measuring the ph value, Measuring the ph value – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual

Page 29

Advertising
background image

Multi 3500i

Operation

27

4.4.2

Measuring the pH value

AutoRead

(Drift control)

The AutoRead function (drift control) continually checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The display of the measured vari-
able flashes until a stable measured value is available.

Criteria

With identical measurement conditions, the following applies:

1

Perform the preparatory activities according to section 4.6.1.

2

Immerse the pH electrode in the test sample.

3

Select the pH or mV display with <M>.

pH

6.949

24.8

°C

Measured
variable

Reproducibility

Response time

pH value

Better than 0.01

> 30 seconds