2 measuring, Measuring – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual

Page 61

Advertising
background image

Multi 3500i

Operation

59

4.7.2

Measuring

You can carry out conductivity measurements as follows:

Selecting the

displayed

measured variable

You can switch between the following displays with <M>:

z Conductivity [

µS/cm] / [mS/cm]

z Specific resistance [k

·cm] / [M

·cm]

z Salinity SaL [ ]
z Total dissolved solids TDS [mg/l]

The factor to calculate the total dissolved solids is set to 1.00 in the fac-
tory. You can adjust this factor to meet your requirements in the range
of 0.40 to 1.00. The factor can be set in the TDS measuring menu.

AutoRead criterion

The measuring instrument checks the stability of the measured value
on the basis of the temperature measurement. For identical measure-
ment conditions, the following applies:

1

Perform the preparatory activities according to section 4.7.1.

2

Immerse the conductivity measuring cell in the test sample.

731

µS/cm

24.8

°C

Measured
variable

Reproducibility

Response time

Temperature

better than 0.02 °C

> 10 seconds