Preparatory activities

Perform the following preparatory activities when you want to measure:

Note
While no valid calibration is available, e.g. in the delivery condition, "Er-
ror" appears in the measured value display.

4.5.2

Measuring the ion concentration

AutoRead

(drift control)

The AutoRead function (drift control) continually checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The display of the measured vari-
able flashes until a stable measured value is available.

Criteria

With identical measurement conditions, the following applies:

Temperature while cali-

brating and measuring

For precise ISE measurements the temperature difference between
measurement and calibration should not be greater that 2 K. Therefore,

1

Connect the ISE combination electrode to the measuring in-
strument. The pH/ISE measuring window is displayed.

2

If necessary, select the ISE display (unit, mg/l) with <M>.

3

Measure the temperature of the test sample using a thermom-
eter.

4

Calibrate or check the measuring instrument with the elec-
trode.

1

Perform the preparatory activities according to section 4.5.1.

2

Immerse the electrode in the test sample.

ISE

0.157

mg/l

25

°C

Measuring signal

Reproducibility

Response time

Electrode
voltage

better than 0.1 mV

> 30 seconds