4 settings for conductivity measuring cells, Settings for conductivity measuring cells – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual

Page 63

Advertising
background image

Multi 3500i

Operation

61

4.7.4

Settings for conductivity measuring cells

Overview

The following settings are possible for conductivity measuring cells:

z Measured variable
z Reference temperature
z Temperature compensation
z TDS factor
z Calibration interval
z Measuring cell/cell constant

Settings

The settings are made in the measuring menu of the conductivity mea-
surement. To open the settings, activate the relevant measuring win-
dow in the measured value display and press the <MENU/ENTER> key
shortly. After completing the settings, switch to the measured value dis-
play with <M>.

Menu item

Possible
setting

Description

Calibration / Calibra-
tion interval

1 ... 999 d

Calibration interval for the
measuring cell (in days).
The measuring instrument re-
minds you to calibrate regu-
larly by the flashing sensor
symbol in the measuring win-
dow.

Calibration / Calibra-
tion record

-

Displays the calibration
record of the last calibration.

Reset

-

Resets all sensor settings to
the delivery condition (see
section 4.10.1).

Temp. comp. (TC) /
Reference temp.

20 °C
25 °C

Reference temperature

This setting is only available
when the Conductivity or
Spec. resist. display is set.

Temp. comp. (TC) /
Compensation

nLF
lin
Off

Procedure for temperature
compensation (see
section 4.7.3).

This setting is only available
when the Conductivity or
Spec. resist. display is set.