Calibration via a

Comparison meas.

For this calibration procedure, the Comparison meas. setting must be
set to On in the measuring menu.

Note
Before calibrating via a comparison measurement, the sensor should
be calibrated in the air calibration vessel.

Proceed as follows to calibrate the instrument:

8

You have the following options:
z Accept the new calibration values with <MENU/ENTER>.

Subsequently, the calibration record is displayed and output
to the interface at the same time.

z To switch to the measured value display without accepting

the new calibration values, press <M> or <ESC>.

i Relative slope = 0.98
i Sensor+++

Accept

ConOx <CAL>

1

Connect the DO sensor to the measuring instrument.

2

Immerse the DO sensor in the reference solution.

3

In the measured value display, select the DO measuring win-
dow with <

> <> and <M>.

4

Start the calibration with <CAL>.
The calibration display appears.

i Immerse sensor in ref. sol.

Continue

ConOx <CAL>