5 displaying and outputting calibration records, Displaying and outputting calibration records – Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual
Page 76

Operation
Multi 3500i
74
4.8.5
Displaying and outputting calibration records
You can display the calibration data
z of one selected sensor, or
z of all sensors connected
and then output them to the interface.
Displaying the calibra-
tion record of a selected
sensor
The calibration record of the last calibration can be found under the Cal-
ibration / Calibration record menu item in the respective measuring
menu. To open the settings, activate the relevant measuring window in
the measured value display and press the <MENU/ENTER> key
shortly.
Displaying the calibra-
tion records of all
sensors connected
The calibration records of the last calibration can be found under the
Data storage / Calibration data storage menu item in the system menu.
To open the system menu in the measured value display, press the
<MENU/ENTER> key for approx. 1 s.
Menu item
Setting/
function
Description
Data storage / Cali-
bration data storage /
Display
-
Displays in pages all cali-
bration records of all sen-
sors.
Further options:
z Scroll through the cali-
bration records with
<
▲> <▼>.
z Output the displayed
calibration record to the
interface with <PRT>.
z Quit the display with
<ESC> or
<MENU/ENTER>.
z Switch directly to the
measured value display
with <M>.
Data storage / Cali-
bration data storage /
Output to RS232
-
Outputs to the interface the
calibration records of all
sensors.