Xylem MULTI 3500i HANDHELD MULTIPARAMETER METERS User Manual

Page 66

Advertising
background image

Operation

Multi 3500i

64

Calibration evaluation

After the calibration, the measuring instrument automatically evaluates
the current status of the calibration. The evaluation appears on the dis-
play and in the calibration record.

Determining the

cell constant

For this calibration procedure, the Measuring cell setting must be set to
cal in the measuring menu. Proceed as follows to determine the cell
constant:

Display

Calibration
record

Cell constant [cm

-1

]

+++

within the ranges
0.450 ... 0.500 cm

-1

or
0.800 ... 1.200 cm

-1

Error

Error

outside the ranges
0.450 ... 0.500 cm

-1

or
0.800 ... 1.200 cm

-1

Perform error elimination ac-
cording to chapter 6 W

HAT

TO

DO

IF

...

1

Connect a conductivity measuring cell to the measuring instru-
ment.

2

In the measured value display, select the conductivity measur-
ing window with <

> <> and <M>.

3

Start the calibration with <CAL>.
The calibration display appears.

4

Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.

i Cell constant = 0.475 1/cm
i Immerse sensor in solution

Continue

<CAL>